19

Removing and Replacing the Battery Pack

Batt ery Replacement Notes:
• To retain history, ensure the pump has been allowed to go to Sleep

aft er the last run.

• Turn off the pump before removing the batt ery. Removing the batt ery

while the pump is on or running may corrupt pump history.

• Programs should be reloaded using DataTrac for Leland Legacy

Soft ware aft er replacing the batt ery pack.

• Sampling time, delayed start, and other sett ings entered using the

pump keypad should be reprogrammed aft er replacing the batt ery
pack.


1. Position pump with belt clip facing upward.

2.

Use a Phillips head screwdriver to remove three screws on bott om half of
pump.

3. Grasp and remove batt ery pack by pulling it up and away from pump

body.

4. Align connector of new batt ery pack with connector in pump body.

5. Gently press new batt ery pack into pump body until it is fl ush with the

pump case and replace the three screws.

Ensure that the long screw is replaced in the top screw hole. Do not
overtighten screws.

For more information on SKC pump batt ery packs,

go to www.skcinc.com/instructions/1756.pdf.

Belt clip

Battery

pack

Battery

connector

Battery Replacement

Screws

(top screw is

longer than

bottom