Pump setup, Resetting run time data, Setting a sampling time (st) – SKC 100-3000 Leland Legacy User Manual
Page 11: Setting a datatrac program
8
10. ESC: Press [] to exit Setup without saving new
sett ings.
11. End: Press [] to save new sett ings and exit Setup.
PrOFF: Appears only when a program is loaded into pump memory.
See DataTrac for Leland Legacy Soft ware Operating Instructions
(Form 40085, included on soft ware CD) for sett ing a program.
See page 18 for Deleting a DataTrac Program or Delayed Start.
Resetting Run Time Data
To reset accumulated volume and run time data to zero:
1. Press
[], then press the security code in
sequence. Setup will display briefl y.
2. Press
until Clr appears, then press [].
3. Press
until End appears, then press [] to exit
Setup. The accumulated data is cleared and the pump
is now in Hold.
CLr does not clear previously set sampling time
(ST). See Deleting a Sampling Time on page 18.
Setting a Sampling Time (ST)
Program the Leland Legacy from the integral keypad or a PC using DataTrac
soft ware to sample from 1 to 99,999 minutes.
1. Press
[], then press the security code in
sequence. Setup will display briefl y.
2. Repeatedly press until ST L/min and a fl ashing time
and Set appear on the display.
3. Set the sampling time by pressing or to increase
or decrease it to the desired time in minutes.
4. Press repeatedly until End appears.
5. Press [] to save the new sampling time and exit
Setup.
6. Press [] to begin sampling. The time display will
count down in minutes and the pump will go to Hold.
The total sampling time will display.
7. To delete a set sampling time, see Deleting a Sampling
Time on page 18.
Setting a DataTrac Program
See DataTrac for Leland Legacy Software Operating Instructions (included on
software CD).
Pump Setup