10. ESC: Press [] to exit Setup without saving new

sett ings.

11. End: Press [] to save new sett ings and exit Setup.

PrOFF: Appears only when a program is loaded into pump memory.

See DataTrac for Leland Legacy Soft ware Operating Instructions
(Form 40085, included on soft ware CD) for sett ing a program.
See page 18 for Deleting a DataTrac Program or Delayed Start.

Resetting Run Time Data

To reset accumulated volume and run time data to zero:
1. Press

[], then press the security code  in

sequence. Setup will display briefl y.

2. Press

 until Clr appears, then press [].

3. Press

 until End appears, then press [] to exit

Setup. The accumulated data is cleared and the pump
is now in Hold.

CLr does not clear previously set sampling time
(ST). See Deleting a Sampling Time on page 18.

Setting a Sampling Time (ST)

Program the Leland Legacy from the integral keypad or a PC using DataTrac
soft ware to sample from 1 to 99,999 minutes.

1. Press

[], then press the security code  in

sequence. Setup will display briefl y.

2. Repeatedly press  until ST L/min and a fl ashing time

and Set appear on the display.

3. Set the sampling time by pressing  or  to increase

or decrease it to the desired time in minutes.

4. Press  repeatedly until End appears.
5. Press [] to save the new sampling time and exit

Setup.

6. Press [] to begin sampling. The time display will

count down in minutes and the pump will go to Hold.
The total sampling time will display.

7. To delete a set sampling time, see Deleting a Sampling

Time on page 18.

Setting a DataTrac Program

See DataTrac for Leland Legacy Software Operating Instructions (included on
software CD).

Pump Setup