Calibration – SKC 100-3000 Leland Legacy User Manual

Page 14

background image

11

Verifying Flow Rate Using a Primary Standard Calibrator

Before use, allow pump to equilibrate after moving it from one

temperature extreme to another.

1. Ensure the pump has run for 5 minutes before performing calibration.
2. Connect the pump inlet to a calibrator with representative media in line

(see photo below).

3. Press [], then press the security code  in sequence. The fl ow

rate and Set will fl ash.

4. Set

the

fl ow on the pump display by pressing  or 

to increase or decrease fl ow to the desired rate.

5. Press . Adj will appear.
6. If the calibrator reads a higher flow rate than the

pump is set for,
press  until they
are in agreement
( w i t h i n 1 0 m l ) .
If the calibrator
reads a lower fl ow
rate, press  until
they agree (within
1 0 m l ) . W h e n
pressing  or  ,
the pump display
will indicate the
adjustment (or correction) made in L/min.

7. Press  until End appears.
8. Press [] to save new fl ow rate and Adj and exit Setup. Reset run

time data (see page 8).

If the pump has been programmed with DataTrac Soft ware and
switched to manual operation, a program may remain in pump
memory. Prog will display in the upper left corner of the pump
display. See page 18 for Deleting a DataTrac Program or Delayed Start.

Calibration train with sample

medium in line

Calibration