Calibration – SKC 100-3000 Leland Legacy User Manual
Page 14
![background image](/manuals/349567/14/background.png)
11
Verifying Flow Rate Using a Primary Standard Calibrator
Before use, allow pump to equilibrate after moving it from one
temperature extreme to another.
1. Ensure the pump has run for 5 minutes before performing calibration.
2. Connect the pump inlet to a calibrator with representative media in line
(see photo below).
3. Press [], then press the security code in sequence. The fl ow
rate and Set will fl ash.
4. Set
the
fl ow on the pump display by pressing or
to increase or decrease fl ow to the desired rate.
5. Press . Adj will appear.
6. If the calibrator reads a higher flow rate than the
pump is set for,
press until they
are in agreement
( w i t h i n 1 0 m l ) .
If the calibrator
reads a lower fl ow
rate, press until
they agree (within
1 0 m l ) . W h e n
pressing or ,
the pump display
will indicate the
adjustment (or correction) made in L/min.
7. Press until End appears.
8. Press [] to save new fl ow rate and Adj and exit Setup. Reset run
time data (see page 8).
If the pump has been programmed with DataTrac Soft ware and
switched to manual operation, a program may remain in pump
memory. Prog will display in the upper left corner of the pump
display. See page 18 for Deleting a DataTrac Program or Delayed Start.
Calibration train with sample
medium in line
Calibration