Pump setup – SKC 100-3000 Leland Legacy User Manual
Page 10
![background image](/manuals/349567/10/background.png)
7
If changing other parameters, do not press [] but continue
pressing aft er End appears and the remainder of the menu
items will appear. Once all changes are entered, press until End
appears, then press [] to save new sett ings and exit Setup.
Pressing [] when Esc appears will exit Setup without saving
new sett ings.
3. CALCh: Use for CalChek calibration feature only.
Pressing [] initiates single-point calibration.
Pressing seven times initiates a full calibration.
See CalChek Calibration instructions on pages 12 to 16.
4. 12 Hr/24 Hr Clock and Delayed Start (factory
default is 12 Hr clock): Press or to move
between standard (12 hour), military (24 hour), and
Dela (delayed start). Press to select. If Dela (delayed
start) is selected, follow instructions on page 9.
5. Time of day: Press or to increase or decrease
fl ashing hour. Press to move from hours to
minutes. Press or to increase or decrease
fl ashing minutes. Press to move to next parameter.
6. ST (Sampling Time): Allows the user to program
a specifi c run time. Press or to increase or
decrease the time in minutes (up to 99,999 minutes).
Press to move to next parameter. See pages 8 and 18
for Sett ing and Deleting a Sampling Time.
7. Temperature (factory default is Celsius): Press
or to toggle between Fahrenheit (F) and Celsius
(C). Press to move to next parameter.
8. Atmospheric
Pressure
(factory default is mm):
Press or to toggle between inches of mercury
(In), millibars (mb), and millimeters of mercury
(mm). Press to move to next parameter.
9. CLr: Press [] to reset accumulated run time and
volume data to zero (see Resett ing Run Time Data on page 8).
Pump Setup