Pump setup, Setting a delayed start – SKC 100-3000 Leland Legacy User Manual

Page 12

background image

9

Setting a Delayed Start

A delayed start can be programmed using the pump
keypad or from a PC using DataTrac Soft ware. The
following instructions are for keypad only. See DataTrac
for Leland Legacy Operating Instructions (included on soft ware
CD) for programming from a PC.

When sett ing the pump for sampling from 1 to 99,999
minutes to begin within the next 12-hour period, follow this
procedure:

1. Press

[], then press the security code  in

sequence. Setup will display briefl y.

2. Press

 until the display reaches the 12 Hr/24 Hr clock.

If delayed start is already programmed, the display will show Dela
(delayed start) in place of 12 Hr. If no delay is programmed, press  or
 until the display shows a fl ashing Dela (delayed start).

3. Press

 until the time of day (fl ashing hours) displays. Select the hour

(time of day) that the pump is to begin sampling (within the next 12
hours) by pressing  or  until the desired hour displays. Press  and
the minutes will fl ash. Press  or  until the desired minutes display.

The time of day entered will be the next occurrence of this time
within the next 12-hour period aft er the delayed start is entered.
There is no a.m. or p.m. designation.

4. Press

 until the ST displays. Press  or  to set the desired run time

in minutes. A delayed start cannot be run unless a sampling time (ST)
is programmed.

5. Press

 until End appears.

6. Press

[] to save sett ings and exit Setup.

7. Prog and a fl ashing Hold will appear in the upper

left corner of the display. The pump is now set for
delayed start.

Once a program is set in the pump, the pump cannot be run
manually. To return to manual pump operation, let the program
run its course or delete the program (see Deleting a DataTrac Program
or Delayed Start on page 18).

Display for

delayed start

Pump Setup