Pump setup, Setting pump flow rate – SKC 100-3000 Leland Legacy User Manual
Page 13
10
Setting Pump Flow Rate
1. Press [], then press the security code in sequence.
2. The fl ow rate and Set will fl ash on the LCD. Press to increase fl ow
rate. Press to decrease fl ow rate. The pump will run while fl ow is set.
3. Once the desired fl ow rate is displayed, press until End appears on
the display. The pump will stop running.
4. Press [] to save the new fl ow rate and exit Setup.
Pump Setup
Flow Rate and Volume Display
Flow Rate displayed on the pump LCD is the fl ow to which the pump has been
calibrated. To maintain fl ow as displayed, the pump automatically adjusts fl ow
during sampling for changes in temperature and atmospheric pressure that may
differ from the temperature and atmospheric pressure present at the time of
calibration. The fl ow rate display does not change from the calibrated fl ow rate.
The pump will fault if it is unable to maintain the calibrated fl ow rate.
Volume displayed on the pump LCD is “corrected” in that it is the result of a
continual calculation of corrected fl ow rate multiplied by sample time. Volume
does not display after 99,999 liters. See Volume Display on page 18.