beautypg.com

Ransburg, Procedures – Ransburg Evolver Solvent Robot Applicator 79190-XXXXXXXX User Manual

Page 30

background image

2. Turn the MicroPak Control Unit

ON! Energize

high voltage.

Shut-Down (Or at Shift End)

1. Turn the MicroPak Control Unit OFF.

2. Flush the lines and allow the solvent to remain

in the lines. See "Procedures"

in the "Mainte-

nance" section.

3. Wipe the gun and robot wrist with a cloth and

a suitable, clean non-polar solvent.

Weekly

1. Check the entire system for damage, leaks

and paint accumulation.

2. Clean the atomizer assembly.

3. Conduct a current output test. See "Proce-

dures" in the "Maintenance" section.

PROCEDURES

Gun Cleaning/Service

(See Figure 8)

1. Flush the paint supply line and the gun paint

passages using a solvent which is compatible with

the material being sprayed. Continue to flush until

all traces of paint are gone.

2. Turn off the solvent supply, actuate paint push-

out air at color changer and trigger the gun. Allow

all of the fluid to drain from the spray gun fluid

passages.

3. Clean the exterior surfaces of the spray gun

with a solvent soaked rag. As long as the gun is

intact, a

polar solvent may be used for all cleaning,

however, after cleaning, wipe off all surfaces with

a

non-polar solvent to reduce conductive residue

on the gun's surface. (See Figure 7 regarding

proper solvent selection.)

>

Failure to use a non-polar solvent will

cause a decrease in voltage at the tip of the

gun. This will significantly decrease transfer

efficiency.

C A U T I O N

!

4. Remove end cap [1]. Removing the end cap

releases tension on all internal spray head compo-

nents. Remove needle spring [2] and valve spring

[3], which are loose after removing the piston cap.

5. Remove the air cap retainer [36] and air cap

[35]. Soak in a non-polar solvent if necessary. If

paint remains in the air cap holes, clean with a

toothpick or similar soft wood object. Air caps are

best cleaned in an ultrasonic cleaner.

6. Remove the air cap locator [33] and fluid tip

[34]. Clean using a non-polar solvent.

7. Tightly grip the plastic needle [6] and unscrew

counter-clockwise to remove the front electrode

needle assembly. A short piece of H-2339 tub-

ing (1/4" OD x 0.175" ID) pressed over the front

needle will assist in unscrewing the assembly.

If required, use needle nose pliers with masking

tape or duct tape. Carefully clean with a non-polar

solvent. Replace any parts that show signs of

wear or damage.

>

Never attempt to clean the air cap holes

with a wire or other metal object. Doing so

may damage the air cap, resulting in distor-

tion of the spray pattern.

C A U T I O N

!

>

If using needle nose pliers to unscrew

the front needle, be very careful. Do not grip

on the tapered sealing surface. If the pliers

slip, they could damage the tapered sealing

surface of the needle.

C A U T I O N

!

Evolver Solventborne Robotic Atomizers - Maintenance

26

Ransburg

AA-03-02.12