Section 4: Calibration
08/13
27
51219 -- Rev. 6
6. Press the
key to select the digit and the
key to increase
the value. A
indicator is displayed on the lower right corner of the
screen. The first digit on the left can only be set to 0 or 1. When the
value on the display is the same as the weight on the platform, press
the
key to store it. The display will indicate
and after two
seconds it will advance to the
menu. Save the calibration
changes by selecting
.
4.2.
– Save Settings Menu
In the
menu, press
key to toggle among the
, and
menus. Press
key to select the setting.
1.
– Save Instrument Settings
Select
to save or backup your instrument settings or changes.
2.
- Load Instrument Settings
Select
to load your previous/ current saved Instrument
configuration and calibration settings into the instrument.
3.
-
Load Instrument Settings
Select
to load the factory default Instrument settings into the
instrument.
! – IMPORTANT -- !
This will change the instrument’s calibration and configuration settings
.
! – IMPORTANT -- !
Remember to re-install JP3 to prevent accidental access to the calibration
upon completing the calibration procedure. Legal for Trade applications
require this jumper to be in place. If the JP3 jumper is not in place, it will not
print
.