Section 4: Calibration

08/13

27

51219 -- Rev. 6

6. Press the

key to select the digit and the

key to increase

the value. A

indicator is displayed on the lower right corner of the

screen. The first digit on the left can only be set to 0 or 1. When the
value on the display is the same as the weight on the platform, press

the

key to store it. The display will indicate

and after two

seconds it will advance to the

menu. Save the calibration

changes by selecting

.


4.2.

– Save Settings Menu

In the

menu, press

key to toggle among the

, and

menus. Press

key to select the setting.

1.

– Save Instrument Settings

Select

to save or backup your instrument settings or changes.

2.

- Load Instrument Settings

Select

to load your previous/ current saved Instrument

configuration and calibration settings into the instrument.

3.

-

Load Instrument Settings

Select

to load the factory default Instrument settings into the

instrument.

! – IMPORTANT -- !

This will change the instrument’s calibration and configuration settings

.

! – IMPORTANT -- !

Remember to re-install JP3 to prevent accidental access to the calibration
upon completing the calibration procedure. Legal for Trade applications
require this jumper to be in place. If the JP3 jumper is not in place, it will not
print

.