Model 9110T NOx Analyzer

Getting Started

Teledyne Analytical Instruments

72

Zero/Span

Open to ZERO AIR inlet

3 2

ZERO CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

Zero/Span

Open to ZERO AIR inlet

3 2

SPAN CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

Zero/Span

Open to SPAN inlet

1 2

3.3.2.6. Gas Conditioner Options

AMMONIA REMOVAL SAMPLE CONDITIONER (OPT 86A)

The 9110T includes a Nafion

®

permeation gas exchange tube to remove H

2

O a from the ozone generator

supply gas stream to a dew point of about -20° C (~600 ppm H

2

O) and effectively remove concentrations

of ammonia (NH

3

) up to about 1 ppm.

An additional Sample Conditioner can be added to the 9110T’s sample gas stream.

SAMPLE

PRESSURE

SENSOR

VACUUM

PRESSURE

SENSOR

O

3

FLOW

SENSOR

PMT

PERMAPU

RE

DRYE

R

EXH

AUST MANIFO

LD

NO/NO

X

VALVE

AUTOZERO

VALVE

Sample Gas

NH

3

Dryer/Scrubber

Figure 3-28:

Pneumatics for Sample Conditioner OPT 86A

ZERO AIR SCRUBBER (OPT 86C), FOR Z/S VALVES