Model 9110T NOx Analyzer

Data Acquisition System (DAS) and APICOM

Teledyne Analytical Instruments

180

7.1.3.9. The Compact Report Feature

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting
each parameter in one channel on a separate line, up to five parameters are reported in one line.

The COMPACT DATA REPORT generally cannot be accessed from the standard DAS front panel
menu, but is available via the instrument’s communication ports by using APICOM or the analyzer’s
standard serial data protocol.

7.1.3.10. The Starting Date Feature

This option allows the user to specify a starting date for any given channel when the user wants to start
data acquisition only after a certain time and date. If the STARTING DATE is in the past (the default
condition), the DAS ignores this setting and begins recording data as defined by the REPORT PERIOD
setting.

The STARTING DATE generally cannot be accessed from the standard DAS front panel menu, but is
available via the instrument’s communication ports by using APICOM or the analyzer’s standard serial
data protocol.

7.1.3.11. Disabling/Enabling Data Channels

Data channels can be temporarily disabled, which can reduce the read/write wear on the Disk-on-Module
(DOM).

To disable a data channel, go to the DAS>EDIT menu as shown in Section 7.1.3 then continue as
follows: