Model 9110T NOx Analyzer
Principles of Operation
Teledyne Analytical Instruments
347
Table 13-4: Relay PCA Status LED’s
LED Color
Function
Status When Lit
(Energized State)
Status When Unlit
(Default State)
D1
Red Watchdog
Circuit
Cycles ON/OFF every 3 Seconds
under direct control of the analyzer’s CPU.
D2
Yellow
Reaction Cell Heater
Heating
Not Heating
D3 Yellow
NO
2
NO Converter Heater
Heating
Not Heating
D4
SPARE
D5
1
Yellow
Internal Span Gas Generator
Perm Tube Oven Heater
Heating Not
Heating
D6
SPARE
D7 Green
Zero/Span
Valve Valve OPEN to span gas flow
Valve OPEN to zero gas flow
D8 Green Sample/Cal
Valve
Valve OPEN to
calibration gas flow
Valve OPEN to
sample gas flow
D9
Green
Auto Zero Valve
Sample gas flow BYPASSES
the reaction cell
Sample gas flow is routed
THROUGH the reaction cell
D10 Green
NO/NO
x
Valve
Gas flow routed THROUGH
NO
2
NO converter
Gas Flow BYPASSES
NO
2
NO converter
D11
2
Green
Dual Span Gas
Select Valve
Valve OPEN to SPAN 1
gas inlet
Valve OPEN to SPAN2 inlet
D12
3
Green
Pressurized Span
Shutoff Valve
Span gas flow SHUTOFF
Span gas flow OPEN
D13
4
Green
Pressurized Zero
Shutoff Valve
Zero gas flow SHUTOFF
Zero gas flow OPEN
D14 - 16
SPARE
1
Only active when the optional internal span gas generator is installed.
2
Only active when the dual pressurized span option is installed.
3
Only active when one of the pressurized span gas options is installed.
4
Only active when one the pressurized zero gas option is installed.