Model 9110T NOx Analyzer

Instrument Maintenance

Teledyne Analytical Instruments

245

Figure 11-3:

Ozone Cleanser Assembly

3. Remove the integrated screws with a Phillips screw driver and remove the scrubber

manifold from the chassis.

4. Using a 9/16” wrench, remove both fittings from the cartridge.

5. Discard the glass wool.

6. Pour the contents of the scrubber manifold onto a sheet of white paper. If

necessary, remove the plug to ensure that all the contents are poured out.

 Notice any discoloration of the contents, which is usually white and slightly

transparent.

 The amount of discolored chemical (usually with yellow tint) may give you an

indication of the lifetime of the chemical in your application.