Min 2, Fv t, No f x ] [no = f – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual
Page 226

Model 9110T NOx Analyzer
EPA Protocol Calibration
Teledyne Analytical Instruments
226
2. Determine the volume (cm3) of the calibrator reaction cell (VRC). If the actual
volume is not known, estimate the volume by measuring the approximate
dimensions of the chamber and using an appropriate formula.
3. Determine the required minimum total flow output (FT) :
Equation
10-2
F
T
= analyzer flow demand (cm
3
/min) x 110/100
If more than one analyzer is to be calibrated at the same time, multiply FT by
the number of analyzers.
4. Calculate the NO concentration [NO]
OUT
needed to approximate 90% of the URL of
the NO
2
analyzer to be calibrated:
Equation
10-3
[NO]
OUT
= URL of analyzer (ppm) x 90/100
5. Calculate the NO flow (F
NO
) required to generate the NO concentration [NO]
OUT
:
Equation 10-4
]
[NO
F
x
]
[NO
=
F
STD
T
OUT
NO
6. Calculate the required flow through the ozone generator (F
O
):
Equation 10-5
F
-
-
ppm
2.75
V
X
F
X
]
[NO
=
F
NO
RC
NO
STD
o
min
7. Verify that the residence time (t
R
) in the reaction cell is <2 min:
Equation 10-6
min
2
NO
O
RC
R
F
F
V
t
8. Verify that the dynamic parameter specification (P
R
) of the calibrator's reaction cell
is >2.75 ppm-min:
- 1220 - Multipoint flammable gas and vapor detection system (50 pages)
- 212R - Thermal conductivity analyzer (28 pages)
- 235 - Thermal conductivity analyzer (38 pages)
- 275R - Portable turbine generator purge gas analyzer (21 pages)
- 2000A-EU - General purpose thermal conductivity analyzer (86 pages)
- 2000XTC - Thermal conductivity analyzer (40 pages)
- 2010A - Split architecture thermal conductivity analyzer (110 pages)
- 2010B - Split architecture thermal conductivity analyzer (98 pages)
- 2020 - Explosion proof thermal conductivity analyzer (80 pages)
- 2120 - Trace Nitrogen in Argon Analyzer (66 pages)
- 2120XL - Trace Nitrogen Analyzer (85 pages)
- 2230R - Process Hydrogen Analyzer (26 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 3rd generation (updated 1/31/11) (30 pages)
- 2240 - Portable Handheld Hydrogen Leak Detector, 3rd generation (revision 2/29/08) (40 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 2nd generation (13 pages)
- 2750 - Portable turbine generator gas analzyer (40 pages)
- 300P - Percent oxygen analyzer (24 pages)
- 306WA - Analog trace oxygen analyzer (46 pages)
- 311 - Portable trace oxygen analyzer (19 pages)
- 311D - Portable trace oxygen analyzer with digital meter (18 pages)
- 311XL - Portable trace oxygen analyzer (18 pages)
- 316RA / RB / RAD / RBD - Oxygen analyzers (24 pages)
- 319R - Oxygen analyzer (23 pages)
- 320 Series - Portable oxygen detectors (24 pages)
- 326, 327 and 328 - Oxygen analyzers (45 pages)
- 329R - Oxygen analyzer (22 pages)
- 335 - Analog control room monitor for personnel safety (24 pages)
- 356WA - Analog trace oxygen analyzer (42 pages)
- 3000MA - Paramagnetic oxygen analyzer (63 pages)
- 3000MA - Paramagnetic oxygen analyzer Addendum (2 pages)
- 3000MB - Paramagnetic oxygen analyzer (59 pages)
- 3000PA - General purpose percent oxygen analyzer (69 pages)
- 3000PAEU - General purpose percent oxygen analyzer (78 pages)
- 3000PB - Bulkhead mount percent oxygen analyzer (82 pages)
- 3000TA - General purpose trace oxygen analyzer (75 pages)
- 3000TA-EU - General purpose trace oxygen analyzer (89 pages)
- 3000TA-XLEU - Trace oxygen analyzer (108 pages)
- 3000TB - Bulkhead mount trace oxygen analyzer (78 pages)
- 3000TB-XL - Trace oxygen analyzer (78 pages)
- 3000ZA - Trace oxygen analyzer (81 pages)
- 3000ZA-3X - Trace oxygen analyzer (72 pages)
- 3000ZA2G - Zirconium oxide analyzer (72 pages)
- 3000 Ultra Trace - PPB oxygen analyzer (72 pages)
- 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range (88 pages)
- 3010MA – Paramagnetic oxygen analyzer, no 0-100% range – (superceded) (88 pages)