beautypg.com

Teledyne 7300a - NDIR analyzer User Manual

Page 84

background image

A-4

Appendix

Model 7300A

Teledyne Analytical Instruments

ATTACHMENT 7300A

Quote "Exceptions" and "GAS PHASE Conditions" for this
application:

1

Response Time is proportional to sample system design for take-off

distance, process pressure, line size, by-pass flow design, dead-

volumes/tee's, sample cell volume and instrument electronics.

2

We assume no particulates that could plug the lines or contaminate the

optical components.

3

Control the pressure in the measuring sample cell to within 0.5 % of the

nominal operating cell pressure (for example: 760 mm ± 4 mm Hg

absolute) is required, return pressure assume to be stable at 1 ATM A.

Maximum cell operation pressure to be 5 psig based upon stan-

dard metal/sapphire cell design.

4

The sample flow through the analyzer should be regulated to within ±

5% of its recommended absolute nominal set point flow value. This will

avoid any pressure drops across any orifices.

5

The temperature of the sample from the take-off thorough the sample

cell must be maintained above its dew point to prevent any moisture or

(vapor pressure) for other possible condensables/mists from occurring

and affecting instrument performance. (Loss of CO2 or other soluble

gases)

6

The sample temperature should not vary at the inlet more than ± 15

O

C

of the calibration temperature, nor exceed the design operating specifi-

cations typically 1-50

O

C.

7

Any background components later found to be present and not speci-

fied in the original application feasibility may void the instrument perfor-

mance should they interfere.

For CO2 the impurity of N2O will interfere. Assume no others at mea-

suring wavelength.

8

We recommend when ambient temperatures exceed outside a 5-45

O

C

range that the instrument and sample system be conditioned appropri-

ately for heating/ cooling without condensation.

9

Should the general purpose or explosion proof instrument be supplied

on a back-panel or as a bulk-mount housing, the customer must install

to protect the instrument to meet harsh ambient conditions for dust,

water, snow, wind, corrosion, etc.