Infrared Gas Analyzer

Operation 4

4-19

Teledyne Analytical Instruments

b. Check for leaks downstream from the sensor, where contamina-

tion may be leaking into the system.

c. Check flowmeter to ensure that the flow is no more than

200SCCM

d. Check temperature controller board.

e. Check gas temperature.

If none of the above as indicated, the sensor may need to be replaced.

Check warranty, and contact Analytical Instruments Customer Service.

4.4.2 Span Cal

The

Span

button on the front panel is used to span calibrate the analyzer.

Span calibration can be performed in either the automatic or manual mode.

CAUTION: If you are spanning the Cal Range by itself (multiple

application analyzers only), use

manual mode

zeroing.

If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use

auto mode

spanning in the Cal Range.

Make sure the span gas is flowing to the instrument.

4.4.2.1

Auto Mode Spanning

Observe all precautions in sections 4.4 and 4.4.2, above. Scroll SPAN and

enter the span function. The screen that appears allows you to select whether the
span calibration is to be performed automatically or manually. Use the UP/
DOWN key to toggle between

AUTO and MAN span settling. Stop when AUTO

appears, blinking, on the display.

Select span

mode: AUTO

Enter

to move to the next screen.

Span Val: 2Ш.ШШ %

To begin span

Use UP/DOWN key to change the span setting value.