Infrared Gas Analyzer

Installation 2

2-9

Teledyne Analytical Instruments

2.3.3.4

Range ID Relays

Four dedicated Range ID relay contacts. For any single application they

are assigned to relays in ascending order. For example: if all ranges have the
same application, then the lowest range is assigned to the Range 1 ID relay,
and the highest range is assigned to the Range 3 ID relay. Range 4 is the Cal
Range ID relay. Table 2-7 lists the pin connections.

Table 2-7: Range ID Relay Connections

Pin

Function

21

Range 1 ID Contact

38

Range 1 ID Contact

22

Range 2 ID Contact

39

Range 2 ID Contact

19

Range 3 ID Contact

18

Range 3 ID Contact

34

Range 4 ID Contact

35

Range 4 ID Contact

2.3.3.5

Network I/O

A serial digital input/output for local network protocol. At this printing,

this port is not yet functional. It is to be used in future options to the instru-
ment. Pins 13 (+) and 29 (–).

2.3.3.6

Remote Valve Connector

The 7300A is a single-chassis instrument, which has no Remote Probe

Unit. Instead, the Remote Valve connector is used as another method for
controlling external sample/zero/span gas valves. See Figure 2-5.

Figure 2-5: Remote Probe Connector Pinouts