beautypg.com

8 trouble shooting chart – Teledyne 7300a - NDIR analyzer User Manual

Page 77

background image

Infrared Gas Analyzer

Maintenance 5

5-7

Teledyne Analytical Instruments

5.8 TROUBLE SHOOTING CHART

SYMPTOM

CORRECTIVE ACTION

A

.

Unit does not turn ON

(a) Check AC voltage at inlet for

when actuating Power

correct voltage. Reset breaker

Switch (Motor does not

or switch to supply power to

turn).

receptacle.

(b) Check power to terminal strip
located inside cabinet. CAUTION:
Terminal strip contains 90-220volt
line voltage which can be hazar-
dous. Check for loose connec-
tion to one of the terminal board
lugs; replace as required.

B. No Zero Control (with

(a) Check that Power is ON.

unit on Zero gas). Out-
put is either dead or
fully saturated.

(b) Check that power supply voltages
are correct (+15VDC, Ä15VDC).

(c) Check output of TP6 on preamp
board with a DC voltmeter (should
be less than + or Ä 3VDC). If
output can be zeroed, check volt-
ages at TPl and TP2 on Linearizer
board (should also be zero). Then
check adjustments on E/I board.

(d) Check that infrared source is
operating. If no light, check
source element resistance and
connections.

(e) Try to execute a manual zero.