2 Installation

Model 7300A

2-8

Teledyne Analytical Instruments

Table 2-6: Remote Calibration Connections

Pin

Function

9

+ Remote Zero

11

– Remote Zero

10

+ Remote Span

12

– Remote Span

40

Cal Contact

41

Cal Contact

Remote Calibration Protocol: To properly time the Digital Remote

Cal Inputs to the Model 7300A Analyzer, the customer's controller must
monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal

Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It

will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact

closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.

For example:

1) Test the CRC. When the CRC is open, Send a zero command

until the CRC closes (The CRC will close quickly.)

2) When the CRC closes, remove the zero command.

3) When CRC opens again, send a span command until the CRC

closes. (The CRC will close quickly.)

4) When the CRC closes, remove the span command.

When CRC opens again, zero and span are done, and the sample is

being analyzed.

Note: The Remote Probe connector provides signals to operate the

zero and span gas valves synchronously. However, if you
have the –C Internal valve option, which includes zero and
span gas inputs, the 7300A automatically regulates the zero,
span and sample gas flow.