Infrared Gas Analyzer

Operation 4

4-29

Teledyne Analytical Instruments

Note: The ranges must be increasing from low to high, for example,

if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then
Range 3 cannot be set to 0–50 % since that makes Range 3
lower than Range 2.

Ranges, alarms, and spans are always set in either percent or ppm units, as

selected by the operator, even though all concentration-data outputs change
from ppm to percent when the concentration is above 9999 ppm.

Note: When performing analysis on a fixed range, if the concentra-

tion rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.

To end the session, send:

st
st

to the analyzer from the computer.

4.8.2 The Curve Algorithm Screen

The Curve Algorithm is a linearization method. It provides from 1 to 9

intermediate points between the ZERO and SPAN values, which can be normal-
ized during calibration, to ensure a straight-line input/output transfer function
through the analyzer. (Before setting the alogorithm curve, each range must
be Zeroed and Spanned
).

Each range is linearized individually, as necessary, since each range will

usually have a totally different linearization requirement.

To linearize the ranges, you must first perform the four steps indicated at

the beginning of section 4.8 Programming. You will then be in the MAIN
MENU and select ALOGORITHM.

4.8.2.1 Checking the linearization

From the

MAIN MENU

screen, select

ALGORITHM, and

Enter

.

Sel rng set algo

—> Ш1 Ш2 Ш3 <—

Use the UP/DOWN key to select the range: 01, 02, or 03. Then press

Enter

.

Gas use: CO2

Range: Ø -10 %
Enter again.