beautypg.com

Appendix a-1 specifications – Teledyne 7300a - NDIR analyzer User Manual

Page 81

background image

A-1

Infrared Gas Analyzer

Appendix

Teledyne Analytical Instruments

Appendix

A-1

Specifications

7300A Digital Control Portion:

Ranges: Three Programmable Ranges, field selectable

within limits (application dependent) and Auto
Ranging

Display: 2 line by 20 alphanumeric VFD accompanied

by 5 digit LED display

Signal Output: 4-20mADC iso or 0-1 VDC negative ground

Alarm: Dual Alarm Relay

Mounting: General Purpose

Operating Temperature: 5-45

o

C (41-113

o

F)

Sample Temperature: 5-45

o

C (41-113

o

F)

Accuracy: ±2% of full scale at constant temperature

Linearity: ±2% of full scale

Repeatability: ±1% of full scale

Drift: ±1% of full scale per week at constant tem-

perature

Response Time: 90% of full scale in less than 5 seconds

Span Stability: Less than ±1% of full scale change per month

Noise: ±1% of full scale

Classification: 7300: Explosion proof, suitable for Class I,

Div 1, Groups B, C, & D

7310: General purpose