Teledyne 7300a - NDIR analyzer User Manual

Page 70

background image

4 Operation

Model 7300A

4-34

Teledyne Analytical Instruments

Select STANDBY to restart the system.

4.9.3 Gain Preset

NOTE: This function will apply only for the analizer that has multiple

range and non-linearity.

For nonlinear application, the signal produced by the infrared detector, will

not correspond to the actual gas concentration. The amplification of each range
will not agree, therefore, the gain must be preset in order for the signal to read
linearly. To set the gain, the following must be performed in sequence.

NOTE:Before setting up this feature, you must have a span gas

containing 90%-100% of the lowest range of the analyzer.

1. Set unit range to lowest range.

NOTE:For output reversal, the lowest range should be range 3, else

the lowest range is range 1.

2. Connect span gas to span inlet.

3. Use UP/DOWN key to scroll to the CAL-INDEPD function.

4. Hold Escape/Enter control to ENTER position for approximately 10

seconds until the upper right connect display “ok” message.

5. Use UP/DOWN scroll to select SPAN function, then setup the

setting to span level. Press ENTER key to span.

NOTE:You must do step #5 before the analyzer return to analyze

mode. If the analyzer returns to analyze function, you must
repeat step 3-5 again.

6. Set range to nex range’

7. Repeat steps 3-5 until you reach the last range of the row.