5 Maintenance

Model 7300A

5-2

Teledyne Analytical Instruments

4. Reassemble Housing as shown in Figure 5-1.

American Fuses

European Fuses

Figure 5-2: Installing Fuses

5.1

Routine Maintenance

The 7300A should be inspected on a regular schedule to be

determined by the maintenance personnel. The system filter and
analyzer measurement cell should be maintained as required by
the quality of the sample.

Periodic calibration of the system will assure accurate analysis of the

sample gas and will also indicate the need for maintenance. A regular
calibration schedule should include recording of the zero and span pot values
for performance trends. Clogging of the sample system and other system
problems will usually show early symptoms during online analysis or during
calibration.

Problems with the system not due to the normal processing of the

sample are sometimes failures which must be corrected.

5.2 Filter

The system filter element (if included) should also be periodically

inspected. Replacement is recommended when the element begins to restrict
sample flow or upon visual inspection looks clogged and dirty.

NOTE: Although not normally included on basic Model 7300A

analyzers, input protection filtering is always recom-
mended. Refer to the system piping schematic for recom-
mended hook-up.