Infrared Gas Analyzer

Installation 2

2-7

Teledyne Analytical Instruments

System Alarm:

Actuates when DC power supplied to circuits is

unacceptable in one or more parameters. Permanently
configured as failsafe and latching. Cannot be de-
feated. Actuates if self test fails.

(Reset by pressing

button to remove power. Then

press

again and any other button EXCEPT

System

to resume.

Further detail can be found in chapter 4, section 4-5.

Table 2-5: Alarm Relay Contact Pins

Pin Contact

45 Threshold Alarm 1, normally closed contact
28 Threshold Alarm 1, moving contact
46 Threshold Alarm 1, normally open contact
42 Threshold Alarm 2, normally closed contact
44 Threshold Alarm 2, moving contact
43 Threshold Alarm 2, normally open contact
36 System Alarm, normally closed contact
20 System Alarm, moving contact
37 System Alarm, normally open contact

2.3.3.3

Digital Remote Cal Inputs

Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra-

tion. (See Remote Calibration Protocol below.) See Table 2-5 for pin
connections.

Zero:

Floating input. A 5 to 24 V pulse input across the + and –
pins puts the analyzer into the

Zero

mode. Either side may be

grounded at the source of the signal. A synchronous signal
must open and close the gas control valves appropriately.

Span:

Floating input. A 5 to 24 V pulse input across the + and –

pins puts the analyzer into the

Span

mode. Either side may

be grounded at the source of the signal. A synchronous signal
must open and close the gas control valves appropriately.

Cal Contact: This relay contact is closed while analyzer is spanning

and/or zeroing. (See Remote Calibration Protocol below.)