0 maintenance – Teledyne 7300a - NDIR analyzer User Manual

Page 71

background image

Infrared Gas Analyzer

Maintenance 5

5-1

Teledyne Analytical Instruments

5.0 Maintenance

Aside from normal cleaning and checking for leaks at the gas connec-

tions, routine maintenance is limited to replacing filter elements and fuses,
and recalibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.0

Replacing the Fuse

Remove Power to Unit before replacing the
fuse.

1. Place small screwdriver in notch, and pry cover off, as shown in

Figure 5-1.

Figure 5-1: Removing Fuse Block from Housing

2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.

3. Replace fuse as shown in Figure 5-2.