Teledyne 7300a - NDIR analyzer User Manual
Page 40
4 Operation
Model 7300A
4-4
Teledyne Analytical Instruments
4.2.1.2 Setup Mode
The MAIN MENU consists of 14 functions you can use to customize and
check the operation of the analyzer. Figure 4-1 shows the functions available
with the 7300A. They are listed here with brief descriptions:
1
AUTO-CAL:
Used to define and/or start an automatic calibration
sequence.
2
PSWD
: Used to establish password protection or change the
existing password.
3
LOGOUT
: Logging out prevents unauthorized tampering with
the analyzer settings.
4
MODEL
: Displays Manufacturer, Model, and Software version
of the instrument.
5
SELF-TEST
: The instrument performs a self-diagnostic routine
to check the integrity of the power supply, output boards, cell and
amplifiers.
6
SPAN
: Set up and/or start a span calibration.
7
ZERO
: start a zero calibration.
8
ALARMS:
Used to set the alarm setpoints and determine
whether each alarm will be active or defeated, HI or LO acting,
latching or not, and failsafe or not.
9
RANGE
: Used to set up three analysis ranges that can be
switched automatically with auto-ranging or used as individual
fixed ranges.
10
HARDW-VAR
: This function displays parameters used by the
software for some calculations. It is a troubleshooting tool to be
used by qualified technicians.
11 FILTER: This function sets the digital filter from 0 to 10, 0
being faster response time, and 10 being the slowest. The default
is 4.
12 CAL_OUTPUT: Calibrate Analog Output
13 APPLICATIONS: Restricted function, not generally accessed
by the end user. Used to define up to three analysis ranges and a
calibration range (including impurity, background low end of
range, high end of range, and % of ppm units).
14 ALOGORITHM: Arestricted function, not generally accessed
by the end user. Used to linearize the output for the range of
interest.