background image

4 Operation

Model 7300A

4-30

Teledyne Analytical Instruments

Algorithm setup:

VERIFY SETUP

UP/DOWN to select and

Enter

VERIFY to check whether the linearization

has been accomplished satisfactorily.

Dpt INPUT OUTPUT

Ш Ш.ШШ Ш.ШШ

The leftmost digit (under

Dpt) is the number of the data point being moni-

tored. Use the UP/DOWN keys to select the successive points.

The

INPUT value is the input to the linearizer. It is the simulated output of

the analyzer. You do not need to actually flow gas.

The

OUTPUT value is the output of the linearizer. It should be the ACTU-

AL concentration of the span gas being simulated.

If the

OUTPUT value shown is not correct, the linearization must be correct-

ed.

ESCAPE

to return to the previous screen. Select and Enter

SET UP to

Calibration Mode screen.

Select algorithm

mode : AUTO

There are two ways to linearize:

AUTO and MANUAL: The auto mode

requires as many calibration gases as there will be correction points along the
curve. The user decides on the number of points, based on the precision re-
quired.

The manual mode only requires entering the values for each correction

point into the microprocessor via the front panel buttons. Again, the number of
points required is determined by the user.

Note: Before performing section 4.8.2 or 4.8.2.3, you must check to

ensure that your calibration gases or points are between low
end and high end of the range setup. All correction points
must be between the Zero and the Span concentration. Do not
enter the Zero and Span points as part of the correction

4.8.2.2 Manual Mode Linearization

To linearize manually, you must have previous knowledge of the nonlinear

thermal-conductivity characteristics of your gases. You enter the value of the
differential between the actual concentration and the apparent concentration
(analyzer output). Analytical Instruments has tabular data of this type for a large
number of gases, which it makes available to customers on request. See Appen-
dix for ordering information. To enter data: