Infrared Gas Analyzer

Installation 2

2-11

Teledyne Analytical Instruments

Table 2-8: Commands via RS-232 Input

Command

Description

as

Immediately starts an autospan.

az

Immediately starts an autozero.

rp

Allows reprogramming of two

System

functions:

APPLICATION (gas use) and ALGORITHM (linearization).

st

Toggling input. Stops/Starts any status message output from
the RS-232, until st is sent again.

Implementation: The RS-232 protocol allows some flexibility in its

implementation. Table 2-9 lists certain RS-232 values that are required by
the Model 7300A implementation.

Table 2-9: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

2.4

Gas Requirements

Instrument Air is required for zeroing of the Infrared Analyzer. It must

be free of oil, particulates and water vapor (that will not condensate,
unfiltered plant air is not recommended).
A supply pressure of 10-50psig
with a typical flow rate of 0.1 to 0.6 SCFH (50-250 cc/min) is needed.
Bottled gas is recommended (air or Nitrogen) if high quality air is not avail-
able.

For accurate calibration, the analyzer requires blended gas mixtures

certified to +/- 2% accuracy.