beautypg.com

Teledyne 7300a - NDIR analyzer User Manual

Page 53

background image

Infrared Gas Analyzer

Operation 4

4-17

Teledyne Analytical Instruments

If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use

auto mode

zeroing in the Cal Range.

Make sure the zero gas is flowing to the instrument. If you get a

CELL

CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.

4.4.1.1

Auto Mode Zeroing

Observe the precautions in sections 4.4 and 4.4.1, above.Scroll to ZERO

function buy using UP/DOWN control and enter the zero function mode. The
screen allows you to select whether the zero calibration is to be performed
automatically or manually. Use the UP/DOWN key to toggle between

AUTO

and

MAN zero settling. Stop when AUTO appears, blinking, on the display.

Select zero

mode: AUTO

Press

Enter

to begin zeroing.

####.## % CO2

Slope=#.### C—Zero

The beginning zero level is shown in the upper left corner of the display. As

the zero reading settles, the screen displays and updates information on

Slope= in

ppm/second (unless the Slope starts within the acceptable zero range and does
not need to settle further). The system first does a course zero, shown in the
lower right corner of the screen as

C—Zero, for approximate 3 min, and then does

a fine zero, and displays

F—Zero, for approximate 3 min.

Then, and whenever Slope is less than 0.01 for at least 3 min, instead of

Slope you will see a countdown:

9 Left, 8 Left, and so fourth. These are software

steps in the zeroing process that the system must complete, AFTER settling,
before it can go back to

Analyze

. Software zero is indicated by

S–Zero in the

lower right corner.

####.## % CO2

4 Left=#.### S—Zero

The zeroing process will automatically conclude when the output is within

the acceptable range for a good zero. Then the analyzer automatically returns to
the

Analyze

mode.