Maintenance – PASCO CI-6729 (1X) CONDUCTIVITY SENSOR User Manual
Page 6
![background image](/manuals/340571/6/background.png)
2
Conductivity Sensor
012–06485B
Maintenance
Cleaning
To ensure accurate and reproducible results, the electrode cell must be clean.
A dirty cell will contaminate the solution and change the conductivity of the
liquid.
Cells can be cleaned with detergent and/or dilute nitric acid (1%) by dipping
or filling the cell with cleaning solution and stirring for three minutes.
Storage
Although the best method to store the electrode is by immersing it in deionized
water, the electrode can be stored dry in its container. If the cell is stored dry,
it should be soaked in distilled water for 5 to 10 minutes before use.
Specifications:
CI-6729 (1X conductivity electrode)
• Range:
20k scale: 0–20,000 µS/cm
2k scale: 0–2,000 µS/cm
200 scale: 0–200 µS/cm
• Conductivity electrode cell constant: 1.0 ± 0.1
• Accuracy:
without calibration < 10%
after calibration < 1%
• Resolution:
20k scale, ± 10 µS
2k scale, ± 1.0 µS
200 scale, ± 0.1 µS
• Temperature range
0–80°C
• Electrode body:
Epoxy
CI-6739A (10X conductivity electrode)
• Range:
20k scale: 0–200,000 µS/cm
2k scale: 0–20,000 µS/cm
200 scale: 0–2000 µS/cm
• Conductivity electrode cell constant: 10 ± 1.0
• Accuracy:
without calibration < 10%
after calibration < 1%
• Resolution:
20k scale, ± 100 µS
2k scale, ± 10 µS
200 scale, ± 1.0 µS
• Temperature range: 0–80°C
• Electrode body:
Epoxy