Conductivity Sensor
012–06485B
➤ Always calibrate the
electrode before measuring
absolute (rather than relative)
concentrations.
Single-Point Calibration
1. Open the Conductivity Sensor calibration dialog box by double-
clicking the Conductivity Sensor icon in the Experiment Setup
window.
2. Make sure that the low value is set at 0.000 µS/cm and 0.000 volts
(default values). If these values are not set at zero, enter 0.000 in
the dialog boxes for Low Value of µS/cm and Volts.
3. Place the Conductivity Sensor into the calibration solution. In the
High Value box type a value in µS/cm that corresponds to the
known weight percent solution. As soon as the voltage reading
stabilizes, click Read to enter the HighValue.
➤ Note: This manual has been written with the assumption that the
user has a basic familiarity with Science Workshop and has access to
the User’s Guide for Science Workshop. Users can gain basic skills by
viewing the training video and by doing the tutorial within Science
Workshop. Another useful resource is the Quick Reference Card for
Science Workshop.