Measurement device technical data

Subject to changes.

Performance specifications

Measurement method

Indophenol blue method, derived from DIN 38406 E5

Measurement ranges

AMTAX inter 2 – 2:
AMTAX inter 2 – 20:
AMTAX inter 2 – 80:

0.02–2.00 mg/L NH

4

-N

0.1–20 mg/L NH

4

-N

2.0–80 mg/L NH

4

-N

Measurement uncertainty

AMTAX inter 2 – 2:
AMTAX inter 2 – 20/80:

±4 % of the measurement value ±0.02 mg/L NH

4

-N

±2 % of the measurement value ±0.02 mg/L NH

4

-N

Measuring interval

5 or 10 min (selectable with continuous sample feed)

Measured value display

Graphic display with curve diagram

Calibration

Automatic at selectable intervals

Inspection interval

6 months

Maintenance requirements Approximately 1 h / week, typical

Reagent stock

Reagents:
Null/standard solution:
Cleaning solution:

4–8 weeks depending on interval
12 months
3–6 months depending on interval

Outputs

Two current outputs:
Two limit-value contacts:
Interface:

0/4...20 mA, maximum 500 Ω
Potential-free, 24 V 1 A
Bus-capable (optional)

Ambient temperature

+5° C to +40° C

Power connector

230 V AC / 50 Hz

Power consumption

Approximately 150 VA

Dimensions: W x H x D

560 mm x 810 mm x 330 mm

Mass

Approximately 29 kg (without reagents)

Fuses

T 2.5 A L; 250 V (2 pieces)

Miscellaneous

Auto-cleaning, data logger, graphic display with curve diagram