Order numbers, Error messages – Hach-Lange AMTAX inter2 Basic User Manual User Manual
Page 120

English 120
Error messages
All notes and error messages show in the bottom left of the display.
Detailed error details can be found in normal cases in the [STATUS]
menu – there, the errors can also be acknowledged.
Order numbers
Further accessories
Error Definition
Measures
SIGNAL LOW
Cuvette not filled
Optics or cuvette
dirty
Lamp failure
Lamp voltage too low
Amplifier defective/not
adjusted
Concentration too high
If neither cleaning nor a lamp
change rectifies the error,
call customer service
SIGNAL HIGH
Cuvette not filled
Lamp voltage too high
Amplifier defective/not
adjusted
Check sample feed
Call customer service
TEST
SETTINGS
Invalid or not plausible
data in the device data
memory
Frequent occurrence
points to an empty
buffer battery
The device data must
be completely checked
(F1 key)
Call customer service
SET
TIME/DATE
Time and/or date have
invalid settings and
were reset to 1/1/93,
15:00
A correction is required
LEAK IN THE
SYSTEM
The moisture sensor
has detected fluid and
after 10 seconds
stopped all pumps
Rectify error, dry out system
and push F1 [MEAS]
(5 seconds) to start the system
again
EPROM ERROR
CALL SERVICE
Error in the
program memory
When this error appears again
after switch-off and another
switch-on, call customer
service
RAM ERROR
CALL SERVICE
Error in the
data storage
When this error appears again
after switch-off and another
switch-on, call customer
service
CALIBRATION
FACTOR
Standard solution or
reagents missing
or are not
OK
Check pumps, hoses and
solenoid valves
Check canister fill level
Item no.
Designation
DOC023.XX.03107
User Manual
Available on request
FILTRAX with 2 m unheated hose
Available on request
FILTRAX with 10 m unheated hose
Available on request
FILTRAX with 20 m unheated hose
Available on request
FILTRAX with 30 m unheated hose
LZX414.00.00000
Tank edge hardware for module carrier
LZX676.00
Control unit bracket
LZX408
VIEWTAX program for data analysis