beautypg.com

Brooks, Model 5861 i – Brooks Instrument 5861i User Manual

Page 8

background image

1-2

Installation and Operation Manual

X-TMF-5861i-MFM-eng

Part Number: 541B111AAG

November, 2008

Brooks

®

Model 5861

i

Section 1 Introduction

Power Requirements
+15 to +28Vdc @ 90 mA
2.5 watts power consumption max.@28 volts

Ambient Temperature Limits
Operating: 5 to 65°C (40 to 150°F)
Non-Operating: -25 to 100°C (-13 to +212°F)

Working Pressure
1500 psi (10.34 MPa) maximum

Output Signal
0-5 Vdc into 2000 ohms or greater. Maximum ripple 3 mV.
Jumper selectable 4-20 mAdc or 0-20 mAdc. Refer to Figure 2-3 for
maximum allowable loop resistance.

Temperature Sensitivity
Zero: Less than ±0.075% F.S. per degree C
Span: Less than ±1.0% F.S. shift over 10-50° C range

Pressure Sensitivity
0.03% per PSI up to 200 PSIG

Power Supply Sensitivity
±0.09% full scale per % power supply voltage variation

Mounting Attitude Sensitivity
±0.5% maximum full scale deviation after re-zeroing

Leak Integrity
1 x 10

-9

Atm. scc/sec Helium

Usable Range
50 to 1

Mechanical Connection
Refer to Figure 2-1.

Electrical Connection
D-type, 15 pin male connector (DA-15P)
Refer to Figure 2-2.