beautypg.com

Brooks, Model 5861 i – Brooks Instrument 5861i User Manual

Page 13

background image

2-5

Installation and Operation Manual
X-TMF-5861i-MFC-eng
Part Number: 541B111AAG
November, 2008

Brooks

®

Model 5861

i

Section 2 Installation

2-7 Electrical Interfacing

To insure proper operation the Model 5861

i must be connected per Figures

2-2, 2-3, and 2-4 and configured according to Section 2-7.

Note: There are several functions on the printed circuit board which are not
used by the Model 5861

i. Avoid making connections marked in Figure 2-2

which are marked *.

The following minimum electrical connections must be made for new
installations:

Chassis Ground
Voltage and/or Current Signal Output
Signal Output Return
+15 to +28 Vdc Supply
Supply Common

Electrical Hook-Up

Signal Output

A 0-5 Vdc flow signal output is always available. The current signal output
is jumper - selectable for either 0-20 mAdc or 4-20 mAdc. The flow signal
output can be measured as a voltage and a current simultaneously on two
different pins of the D-connector. Pin 2 indicates the flowrate with a 0-5
Vdc signal proportional to the mass flow rate. Pin 4 indicates the flowrate
with either a 0-20 mAdc or 4-20 mAdc current signal as determined by the
jumpers on the printed circuit board (Refer to Section 2-7 for jumper
positions). Both the current and voltage signals are returned on Pin 10 of
the D-connector.

Figure 2-3 Maximum Allowable Loop Resistance