Section 4 Maintenance

Table 4-2 Sensor Troubleshooting

Trouble

Possible Cause

Check/Corrective Action

Output stays at 0 Volts

Clogged Sensor.

Clean sensor. Refer to cleaning procedure. Section 4-2.

regardless of flow.

Output signal stays at a voltage Defective PC Board.

Replace PC Board. Refer to Section 4-4.

greater than 5.000 volts and
there is no flow through meter

Defective Sensor.

Replace Sensor Assembly. Refer to Section 4-4.

Meter grossly out of calibration. Partially clogged sensor.

Clean sensor, refer to the cleaning procedure.

Flow is higher than indicated.

Refer to Section 4-2.

Meter grossly out of calibration. Partially clogged restrictor.

Replace restrictor. Refer to Section 4-4.

Flow is lower than indicated.

Meter output oscillates.

Anticipate potentiometer out of adjustment. Adjust anticipate potentiometer. Refer to Section 3-4.

Faulty pressure regulator.

Check regulator output.

Defective PC Board.

Replace PC Board. Refer to Section 4-4.

Output voltage works properly

Loop resistance is too high.

Lower loop resistance.

but output current doesn't change

Loop reading device not isolated.

Use isolated reading device.

Defective PC board

Replace board,Refer to Section 4-4.