Step 5: Verify Parameters and Program the BEI Sensor

1. Click “Verify Parameters” for either Sensor 1 or Sensor 2. This step checks to make sure that

all the needed values have been entered correctly.

If there is an error the software will indicate what

needs to be fixed.

If there is no error the software will grey out the

sensor under program.

2. Click “Begin Programming” to start the programming process, this initializes the chip and

prepares it for programming.





3. If you want to insure correlation between the two Sensors, select “Synchronous

Programming”.

*NOTE: This feature is only available if Sensor 1 and

Sensor 2 have the same “Programming Points”

selected.


4. Position the sensor at the desired location for Point A and click “Program A” without moving

the part.