CEMB USA N500 User Manual
Page 76
A - 2
Specification
-
Measurement specifications
-
A/D converter: resolution 24 bit
-
Dynamic range: > 100 dB
-
Number of averages: from 1 to 16
-
Resolution: 100, 200, 400, 800, 1600, 3200 lines
-
Frequency range: DC – 15kHz max
-
Background noise: typically less than 1.50
µ
V for a spectrum with 400 lines having
a max. frequency of 1 kHz
-
Window: Hanning (always enabled)
-
Speed of analysis: 2.5 averages/sec (400 lines – 1kHz)
-
Data memorization capacity: max 500 spectra and 150 waveforms or monitoring
-
Instrument error limit: 5%