beautypg.com

Calibration sequence – CEMB USA N500 User Manual

Page 48

background image

7 - 6

Balancer mode

Calibration sequence

The calibration operation, necessary for assessing the unbalance of a rotor, is normally a
procedure consisting of various steps. Above all, for the most common two cases, it
consists of:

- Calibration for balancing on one plane:

1) first spin without test weight
2) second spin with test weight on the balancing plane

- Calibration for balancing on two planes:

1) first spin without test weight
2) second spin with test weight only on the first balancing plane
3) third spin with test weight only on the second balancing plane

For the two configurations

correction on one plane with one sensor

correction on two planes with two sensors

The calibration sequence screen on the N500 instrument is organized as in the figures.

2

3

4

7
9

1

5

6

8

10

2

3

4

7

9

1

5

6
8

10