beautypg.com

Agilent e1364a form c switch error messages, Appendix c – Agilent Technologies 75000 SERIES B User Manual

Page 71

background image

Appendix C

Agilent E1364A Form C Switch Error

Messages

Table C-l lists the error messages associated with the Form C Switch
module programmed by SCPI. See the appropriate mainframe manual for a
complete list of error messages.

Table C-1. Form C Switch Error Messages

No.

Title

Potential Cause(s)

-211

Trigger ignored

Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.

-213

Init Ignored

Attempting to execute an INIT command when a scan is already in progress.

-224

Illegal parameter value

Attempting to execute a command with a parameter not applicable to the
command.

1500

External trigger source
already allocated

Assigning an external trigger source to a switchbox when the trigger source
has already been assigned to another switchbox.

2000

Invalid card number

Addressing a module (card) in a switchbox that is not part of the switchbox.

2001

Invalid channel number

Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g., channel 99 of a multiplexer module).

2006

Command not supported on
this card

Sending a command to a module (card) in a switchbox that is unsupported
by the module.

2008

Scan list not initialized

Executing a scan without the INIT command.

2009

Too many channels in
channel list

Attempting to address more channels than available in the switchbox.

2012

Invalid Channel Range

Invalid channel(s) specified in SCAN <channel_list> command. Attempting to
begin scanning when no valid <channel_list> is defined.

2600

Function not supported on
this card

Sending a command to a module (card) in a switchbox that is not supported
by the module or switchbox.

2601

Channel list required

Sending a command requiring a <channel_list> without the <channel_list>.

Appendix C

Agilent E1364A Form C Switch Error Messages 69