ARM

The ARM subsystem selects the number of scanning cycles (1 to 32767) for
each INITiate command.

Subsystem Syntax

ARM
:COUNt

<number> MIN | MAX

:COUNt?

[MIN

|

MAX]

:COUNt

ARM:COUNt <number> MIN | MAX

allows scanning cycles to occur a

multiple of times (1 to 32767) with one INITiate command when
INITiate:CONTinuous OFF | 0 is set. MIN sets 1 cycle, MAX sets 32767
cycles.

Parameters

Parameter Name

Parameter Type

Range of Values

<number>

numeric

1-32767 l MIN l MAX

Comments

Number of Scans: Use only values between 1 and 32767 for the
number of scanning cycles.

Related Commands: ABORt, INITiate[:IMMediate]

*RST Condition: ARM:COUNt 1

Example

Setting Ten Scanning Cycles:
This example sets a Form C Switch for 10 scans of channels 00 through 03.
When the scan sequence completes, channels 00 through 03 (relays 00
through 03) are closed.

ARM COUN 10

Set 10 scans per INIT command

SCAN (@100:103)

Scan channels 00-03

INIT

Start scan, close channel 00

Chapter 5

Agilent E1364A Form C Switch Command Reference 45