beautypg.com

Route, Close – Agilent Technologies 75000 SERIES B User Manual

Page 54

background image

[ROUTe:]

The ROUTe command subsystem controls switching and scanning
operations for Form C Switch modules in a switchbox.

Subsystem Syntax

[ROUTe:]
CLOSe <channel_list
>
CLOSe? <channel_list
>
OPEN <channel_list
>
OPEN? <channel_list
>
SCAN

<channel_list>

:MODE

NONE

|

VOLT

:MODE?

CLOSe

[ROUTe:]CLOSe <channel_list>

closes the Form C Switch channels

specified by channel_list. channel_list has the form (@ccnn) where cc =
card number (01-99) and nn = channel number (00-15).

Parameters

Parameter Name

Parameter Type

Range of Values

<channel_list>

numeric

cc00 - cc15

Comments

Closing Channels: To close a single channel, use [ROUTe:]CLOSe
(@ccnn); for multiple channels use [ROUTe:]CLOSe
(@ccnn,ccnn,); for sequential channels, use [ROUTe:]CLOSe
(@ccnn:ccnn); for groups of sequential channels, use
[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn); or any combination.
Closure order for multiple channels with a single command is not
guaranteed.

Related Commands: [ROUTe:]OPEN, [ROUTe:]CLOSe?

*RST Condition: All channels open.

Example

Closing Form C Switch Channels:
This example closes channels 100 and 213 of a two-module switchbox (card
numbers 01 and 02).

CLOS (@100,213)

Close channels 100 and 213. 100
closes channel 00 of card #1 and
213 closes channel 13 of card #2.

52 Agilent E1364A Form C Switch Command Reference

Chapter 5