INITiate

The INITiate command subsystem selects continuous scanning cycles and
starts the scanning cycle.

Subsystem Syntax

INITiate
:CONTinuous

<mode>

:CONTinuous?
[:IMMediate]

:CONTinuous

INITiate:CONTinuous ON | OFF | 1 | 0

enables or disables continuous

scanning cycles for the switchbox.

Parameters

Parameter Name

Parameter Type

Range of Values

<

mode>

boolean

ON | OFF | 1 | 0

Comments

Continuous Scanning Operation: Continuous scanning is enabled
with the INITiate:CONTinuous ON or INITiate:CONTinuous 1
command. Sending the INITiate[:IMMediate] command closes the
first channel in the channel list. Each trigger from a trigger source
selected by the TRIGger:SOURce command advances the scan
through the channel list. A trigger at the end of the channel list closes
the first channel in the channel list and the scan cycle repeats.

Non-Continuous Scanning Operation: Non-continuous scanning is
enabled with the INITiate:CONTinuous OFF or
INITiate:CONTinuous O command. Sending the
INITiate[:IMMediate] command closes the first channel in the
channel list. Each trigger from a trigger source selected by the
TRIGger:SOURce command advances the scan through the channel
list. At the end of the scanning cycle, the last channel in the channel
list is closed.

Stopping Continuous Scan: See the ABORt command.

Related Commands: ABORt, ARM:COUNt, TRIGger,
TRIGger:SOURce

*RST Condition: INITiate:CONTinuous OFF

Example

Enabling Continuous Scanning:
This example enables continuous scanning of channels 00 through 03 of a
single-module switchbox. Since TRIGger:SOURce IMMediate (default) is
set, use an interface clear command (such as CLEAR) to stop the scan.

INIT:CONT ON

Enable continuous scanning

SCAN (@100:103)

Defines channel list

INIT

Start scan cycle, close channel 00

Chapter 5

Agilent E1364A Form C Switch Command Reference 49