beautypg.com

Scanning channels – Agilent Technologies 75000 SERIES B User Manual

Page 33

background image

Scanning Channels

For the Form C Switch, scanning channels consists of closing a
specified set of channels, one channel at a time. You can scan any
combination of channels for a single-module or a multiple-module
switchbox.

Single, multiple, or continuous scanning modes are available. See
Chapter 4 for information on scanning Form C Switch channels.

Example: Scanning
Using Trig Out Port

This example shows one way to synchronize instrument measurements of
devices under test (DUT) with Form C Switch channel closures. For
measurement synchronization, the Agilent E1300B/E1301B Trig Out BNC
port is connected to the instrument External Trigger In port. See Figure 3-6
for typical user connections.

For this example, the mainframe and instrument are connected via GPIB to
an Agilent E1300B/E1301B mainframe with an address of 709 and an
external instrument (such as an Agilent 3457A Multimeter) with an address
of 722. The Form C Switch is at logical address 120 (secondary address 15).

Note

You must add required instrument commands to line 10. Also, you may
need to add a WAIT statement as line 65, depending on the speed of the
measurement.

10 OUTPUT 722; "TRIG EXT;.. "

! Ext triggering

20 OUTPUT 70915; "OUTP ON"

! Enable Trig Out Port

30 OUTPUT 70915; "TRIG:SOUR BUS"

! Bus triggering

40 OUTPUT 70915; "SCAN (@100:102)"

! Scan channels 00-02

50 OUTPUT 70915; "INIT"

! Enable scan

60 FOR l=1 TO 3

! Start loop

70 ENTER 722:A

! Enter result

80 PRINT A

! Display result

90 TRIGGER 70915

! Advance scan

100 NEXT I

! Increment count

110 END

Chapter 3

Using the Agilent E1364A Form C Switch 31