beautypg.com

Trlgger, Cpon, Immediate – Agilent Technologies 75000 SERIES B User Manual

Page 60

background image

:CPON

SYSTem:CPON <number> | ALL

sets the selected module (card) in a

switchbox to its power-on state.

Parameters

Parameter Name

Parameter Type

Range of Values

<number>

numeric

1 - 99

Comments

Form C Switch Module Power-On State: The power-on state is all
channels (relays) open. Note that *RST opens all channels of all
modules in a switchbox, while SYSTem:CPON <number> opens the
channels in only the module (card) specified in the command.

Example

Setting Card #1 Module to its Power-On State:

SYST:CPON 1

Sets module #1 to power-on state

TRlGger

The TRIGger command subsystem controls the triggering operation of
Form C Switch modules in a switchbox.

Subsystem Syntax

TRIGger
[:IMMediate]
:SOURce BUS | EXTernal | HOLD | IMMediate
:SOURce?

[:IMMediate]

TRIGger[:IMMediate]

causes a trigger event to occur when the defined

trigger source is TRIGger:SOURce BUS or TRIGger:SOURce HOLD.

Comments

Executing the TRIGger[:IMMediate] Command: A channel list
must be defined with [ROUTe:]SCAN <channel_list> and an
INITiate[:IMMediate] command must be executed before
TRIGger[:IMMediate] will execute.

BUS or HOLD Source Remains: If selected, the TRIGger:SOURce
BUS or TRIGger:SOURce HOLD commands remain in effect after
triggering a switchbox with the TRIGger[:IMMediate] command.

Related Commands: INITiate, [ROUTe:]SCAN

58 Agilent E1364A Form C Switch Command Reference

Chapter 5