Trlgger, Cpon, Immediate – Agilent Technologies 75000 SERIES B User Manual
Page 60

:CPON
SYSTem:CPON <number> | ALL
sets the selected module (card) in a
switchbox to its power-on state.
Parameters
Parameter Name
Parameter Type
Range of Values
<number>
numeric
1 - 99
Comments
•
Form C Switch Module Power-On State: The power-on state is all
channels (relays) open. Note that *RST opens all channels of all
modules in a switchbox, while SYSTem:CPON <number> opens the
channels in only the module (card) specified in the command.
Example
Setting Card #1 Module to its Power-On State:
SYST:CPON 1
Sets module #1 to power-on state
TRlGger
The TRIGger command subsystem controls the triggering operation of
Form C Switch modules in a switchbox.
Subsystem Syntax
TRIGger
[:IMMediate]
:SOURce BUS | EXTernal | HOLD | IMMediate
:SOURce?
[:IMMediate]
TRIGger[:IMMediate]
causes a trigger event to occur when the defined
trigger source is TRIGger:SOURce BUS or TRIGger:SOURce HOLD.
Comments
•
Executing the TRIGger[:IMMediate] Command: A channel list
must be defined with [ROUTe:]SCAN <channel_list> and an
INITiate[:IMMediate] command must be executed before
TRIGger[:IMMediate] will execute.
•
BUS or HOLD Source Remains: If selected, the TRIGger:SOURce
BUS or TRIGger:SOURce HOLD commands remain in effect after
triggering a switchbox with the TRIGger[:IMMediate] command.
•
Related Commands: INITiate, [ROUTe:]SCAN
58 Agilent E1364A Form C Switch Command Reference
Chapter 5