Special 488.2 commands – Agilent Technologies 75000 SERIES B User Manual
Page 41
Special 488.2 Commands
This section contains information about the response of the switchbox
instrument to three IEEE 488.2 common commands.
Storing States
The *SAV <numeric_state> command saves the current instrument state.
The state number (0-9) is specified by the <numeric_state> parameter. The
settings saved by this command are:
•
ARM:COUNt
•
TRIGger:SOURce
•
OUTput:STATe
•
INITiate:CONTinuous
•
SCAN (the scan list is set to invalid; therefore, the command does
not save a scan list.)
•
SCAN:MODE
•
SCAN:PORT
Recalling States
The *RCL <numeric_state> command recalls the state when the last *SAV
was executed for the specified <numeric_state> parameter (0-9). If no
*SAV was executed for the <numeric_state>, *RST default settings are
used. Refer to the *SAV settings list for the settings recalled by *RCL.
The *RST command opens all channels, invalidates the current channel list
for scanning and sets the following:
•
ARM:COUNt 1
•
TRIGger:SOURce IMM
•
INIT:CONTinuous OFF
•
OUTPut:STATe OFF
•
SCAN:MODE NONE
•
SCAN PORT NONE
Chapter 4
Understanding the Agilent E1364A Form C Switch 39