Monitor[:state – Agilent Technologies 75000 SERIES B User Manual
Page 50

MONitor[:STATe]
DISPlay:MONitor[:STATe] <mode>
turns the monitor mode on or off.
Parameters
Parameter Name
Parameter Type
Range of Values
<mode>
boolean
ON l OFF | 1 | 0
Comments
•
Monitoring Switchbox Channels: DISPlay:MONitor[:STATe] ON
or DISPlay:MONitor[:STATe] 1 turns the monitor mode ON to show
the channel state of the selected module. DISPlay:MONitor[:STATe]
OFF or DISPlay:MONitor[:STATe] 0 turns the monitor mode OFF.
•
Selecting the Module to be Monitored: Use the
DISPlay:MONitor:CARD <number> |AUTO command to select the
module.
•
Monitor Mode on an Agilent E1301B Mainframe Display: A
typical display for the Agilent E1364A Form C Switch module
follows, where all channels (all relays) are closed.
0,1,2,3,4,5,6,7,8,9,10,11,12,13,14,15
•
*RST Condition:
DISPlay:MONitor[:STATe] OFF | 0
Example
Enabling the Monitor Mode:
DISP:MON:CARD 2
Selects module #2 in a switchbox
DISP:MON 1
Turns monitor mode ON
48 Agilent E1364A Form C Switch Command Reference
Chapter 5