Atec Rohde-Schwarz-ESU-40 User Manual
Page 9
Version 02.00, October 2007
R&S
®
ESU EMI Test Receiver
9
R&S
®
ESU26/40
20 MHz ≤ f < 30 MHz, BW = 9 kHz
<–16 dBµV
30 MHz ≤ f < 500 MHz, BW = 120 kHz
<–5 dBµV
500 MHz ≤ f < 1 GHz, BW = 120 kHz
<–3 dBµV
1 GHz ≤ f < 2 GHz, BW = 1 MHz
<6 dBµV
2 GHz ≤ f < 3 GHz, BW = 1 MHz
<7 dBµV
3 GHz ≤ f ≤ 3.6 GHz, BW = 1 MHz
<11 dBµV
max peak
typ. +11 dB
RMS
typ. +1 dB
quasi peak
band A
typ. +3 dB
band B
typ. +4 dB
Increase of DANL relative to AV display
bands C and D
typ. +6 dB
Immunity to interference
f ≤ 3.6 GHz
>90 dB suppression, typ. >110 dB
f > 3.6 GHz
>70 dB suppression, typ. >100 dB
Image frequency
f = receive frequency
f ≤ 3.6 GHz
>90 dB suppression, typ. >110 dB
3.6 GHz < f ≤ 4.2 GHz
typ. 70 dB suppression
f > 4.2 GHz
>70 dB suppression, typ. >90 dB
Intermediate frequency
f = receive frequency
Spurious response
f > 1 MHz, without input signal,
RF attenuation = 0 dB, termination = 50 Ω
<–103 dBm
∆f > 100 kHz
mixer level <–10 dBm, f
in
≤ 2.3 GHz
<–80 dBc
mixer level <–35 dBm, 2.3 GHz in <4 GHz <–70 dBc mixer level <–10 dBm 4 GHz ≤ f < 8 GHz <–70 dBc 8 GHz ≤ f < 16 GHz <–64 dBc 16 GHz ≤ f < 26 GHz <–58 dBc 26.5 GHz ≤ f < 40 GHz <–52 dBc Other interfering signals f = receive frequency Level display (analyzer mode) 625 × 500 pixels (one diagram), max. Logarithmic level axis 1 dB to 200 dB, in steps of 1/2/5 Linear level axis 10 % of reference level per level division, 1 measurement diagram 3 Number of traces 2 measurement diagrams 6 Trace detector max peak, min peak, auto peak (normal), default value 625 Number of measurement points range 155 to 30001 in steps of about a factor of 2 Trace functions clear/write, max hold, min hold, average local measurement, display update rate, 80 per second remote measurement, display off: 70 per second Trace update rate span = 10 MHz, sweep time 2.5 ms 50 per second Setting range of reference level logarithmic level display –130 dBm to (+5 dBm + RF attenuation), linear level display 7.0 nV to 7.07 V in steps of 1 % logarithmic level display dBm, dBµV, dBmV, dBµA, dBpW Units of level axis linear level display µV, mV, µA, mA, pW, nW
Screen
2 diagrams with independent settings
10 divisions or logarithmic scaling
sample, RMS, average, quasi peak,
CISPR-AV, CISPR-RMS
625 points, zero span
zero span/sweep time 1 ms
max. 30 dBm, in steps of 0.1 dB