19
RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW
preselection = ON, preamplifier = ON
4
f < 10 kHz: RBW = 10 Hz, VBW = 30 Hz, trace average, sample detector,
sweep count = 20, mean marker
f ≥ 10 kHz: RBW = 1 kHz, VBW = 3 kHz, span = 0 Hz, sweep time = 50 ms, trace
average, sample detector, sweep count = 20, mean marker
all models
1 kHz
<–130 dBm
10 kHz
<–140 dBm
100 kHz
<–140 dBm
1 MHz
<–150 dBm
10 MHz
<–165 dBm
R&S
®
ESU8:
20 MHz ≤ f < 500 MHz
<–165 dBm, typ. –168 dBm
500 MHz ≤ f < 2 GHz
<–162 dBm, typ. –165 dBm
2 GHz ≤ f < 3 GHz
<–159 dBm, typ. –162 dBm
3 GHz ≤ f < 3.6 GHz
<–154 dBm, typ. –157 dBm
3.6 GHz ≤ f < 6 GHz
<–162 dBm, typ. –165 dBm
6 GHz ≤ f ≤ 8 GHz
<–160 dBm, typ. –163 dBm
R&S
®
ESU26:
20 MHz ≤ f < 2 GHz
<–162 dBm, typ. –165 dBm
2 GHz ≤ f < 3 GHz
<–158 dBm, typ. –161 dBm
3 GHz ≤ f < 3.6 GHz
<–155 dBm, typ. –158 dBm
3.6 GHz ≤ f < 13 GHz
<–165 dBm, typ. –168 dBm
13 GHz ≤ f < 22 GHz
<–163 dBm, typ. –165 dBm
22 GHz ≤ f ≤ 26.5 GHz
<–160 dBm, typ. –163 dBm
R&S
®
ESU40:
20 MHz ≤ f < 2 GHz
<–162 dBm, typ. –165 dBm
2 GHz ≤ f < 3 GHz
<–158 dBm, typ. –161 dBm
3 GHz ≤ f < 3.6 GHz
<–155 dBm, typ. –158 dBm
3.6 GHz ≤ f < 13 GHz
<–165 dBm, typ. –168 dBm
13 GHz ≤ f < 22 GHz
<–163 dBm, typ. –165 dBm
22 GHz ≤ f < 32 GHz
<–160 dBm, typ. –163 dBm
32 GHz ≤ f ≤ 40 GHz
<–155 dBm, typ. –158 dBm
0 dB RF attenuation, termination 50 Ω, log. scaling, normalized to 1 Hz RBW,
preselection = OFF, preamplifier = OFF, low-noise preamplifier = ON
RBW = 1 kHz, VBW = 3 kHz, zero span, sweep time 50 ms, trace average, sweep
count = 20, mean marker
all models
100 kHz
<–140 dBm
1 MHz
<–150 dBm
10 MHz
<-163 dBm
R&S
®
ESU8
20 MHz ≤ f < 2 GHz
<–165 dBm, typ. –168 dBm
2 GHz ≤ f < 3.6 GHz
<–163 dBm, typ. –166 dBm
3.6 GHz ≤ f < 6 GHz
<–162 dBm, typ. –165 dBm
6 GHz ≤ f ≤ 8 GHz
<–160 dBm, typ. –163 dBm
R&S
®
ESU26:
20 MHz ≤ f < 2 GHz
<–164 dBm, typ. –167 dBm
2 GHz ≤ f < 3 < 3.6 GHz
<–163 dBm, typ. –166 dBm
3.6 GHz ≤ f < 13 GHz
<–165 dBm, typ. –168 dBm
13 GHz ≤ f < 22 GHz
<–163 dBm, typ. –166 dBm
22 GHz ≤ f ≤ 26.5 GHz
<–160 dBm, typ. –163 dBm
R&S
®
ESU40:
20 MHz ≤ f < 2 GHz
<–164 dBm, typ. –167 dBm
2 GHz ≤ f < 3 < 3.6 GHz
<–163 dBm, typ. –166 dBm
3.6 GHz ≤ f < 13 GHz
<–165 dBm, typ. –168 dBm
13 GHz ≤ f < 22 GHz
<–163 dBm, typ. –166 dBm
22 GHz ≤ f < 32 GHz
<–160 dBm, typ. –163 dBm
32 GHz ≤ f ≤ 40 GHz
<–155 dBm, typ. –158 dBm
4
With option R&S
®
ESU-B24 the frequency range of the preamplifier is extended to the upper frequency limit of the instrument.