beautypg.com

Atec Rohde-Schwarz-ESU-40 User Manual

Page 21

background image

Version 02.00, October 2007

R&S

®

ESU EMI Test Receiver

21

R&S

®

ESU26:

20 MHz ≤ f < 30 MHz, BW = 9 kHz

<–15 dBµV

30 MHz ≤ f < 1 GHz, BW = 120 kHz

<–4 dBµV

1 GHz ≤ f < 2 GHz, BW = 1 MHz

<5 dBµV

2 GHz ≤ f < 3 GHz, BW = 1 MHz

<9 dBµV

3 GHz ≤ f < 3.6 GHz, BW = 1 MHz

<12 dBµV

3.6 GHz ≤ f < 13 GHz, BW = 1 MHz

<2 dBµV

13 GHz ≤ f < 22 GHz, BW = 1 MHz

<4 dBµV

22 GHz ≤ f ≤ 26.5 GHz, BW = 1 MHz

<7 dBµV

R&S

®

ESU40:

20 MHz ≤ f < 30 MHz, BW = 9 kHz

<–15 dBµV

30 MHz ≤ f < 1 GHz, BW = 120 kHz

<–4 dBµV

1 GHz ≤ f < 2 GHz, BW = 1 MHz

<5 dBµV

2 GHz ≤ f < 3 GHz, BW = 1 MHz

<9 dBµV

3 GHz ≤ f < 3.6 GHz, BW = 1 MHz

<12 dBµV

3.6 GHz ≤ f < 13 GHz, BW = 1 MHz

<2 dBµV

13 GHz ≤ f < 22 GHz, BW = 1 MHz

<4 dBµV

22 GHz ≤ f < 32 GHz, BW = 1 MHz

<7 dBµV

32 GHz ≤ f ≤ 40 GHz, BW = 1 MHz

<12 dBµV

Level measurement uncertainty

RBW = 10 kHz, level –30 dBm, reference level –30 dBm, RF attenuation 10 dB

preselection = OFF, preamplifier = OFF,
low-noise preamplifier = OFF

6


<0.2 dB (σ = 0.07 dB)

low-noise preamplifier = ON

6

<0.3 dB (σ = 0.1 dB)

Absolute level uncertainty at 128 MHz

preselection/preamplifier = ON

7

<0.3 dB (σ = 0.1 dB)

DC coupling, RF attenuation ≥10 dB, preselection = OFF, preamplifier = OFF,
low-noise preamplifier = OFF

6

+20 °C to +30 °C

20 Hz ≤ f < 10 MHz

<0.5 dB (σ = 0.16 dB)

10 MHz ≤ f < 2 GHz

<0.3 dB (σ = 0.1 dB)

2 GHz ≤ f < 3.6 GHz

<0.5 dB (σ = 0.16 dB)

3.6 GHz ≤ f < 8 GHz, span < 1 GHz

<1.5 dB (σ = 0.5 dB)

8 GHz ≤ f ≤ 40 GHz, span < 1 GHz

<2.0 dB (σ = 0.7 dB)

RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz


add 0.5 dB to above values

+5 °C to +45 °C

20 Hz ≤ f < 3.6 GHz

<0.6 dB (σ = 0.2 dB)

3.6 GHz ≤ f < 8 GHz

add 0.5 dB to above values

8 GHz ≤ f < 26.5 GHz

add 1.0 dB to above values

26.5 GHz ≤ f ≤ 40 GHz

add 1.5 dB to above values

Frequency response
referenced to 128 MHz

RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz


add 0.5 dB to above values

DC coupling, RF attenuation ≥10 dB,
preselection = OFF, preamplifier = OFF, low-noise preamplifier = ON

6

+20 °C to +30 °C

100 kHz ≤ f < 10 MHz

<0.8 dB (σ = 0.26 dB)

10 MHz ≤ f < 3.6 GHz

<0.6 dB (σ = 0.2 dB)

3.6 GHz ≤ f < 8 GHz, span < 1 GHz

<1.5 dB (σ = 0.5 dB)

8 GHz ≤ f ≤ 40 GHz, span < 1 GHz

<2.0 dB (σ = 0.7 dB)

RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz


add 0.5 dB to above values

+5 °C to +45 °C

100 kHz ≤ f < 10 MHz

<1.0 dB (σ = 0.33 dB)

10 MHz ≤ f < 3.6 GHz

<0.8 dB (σ = 0.26 dB)

3.6 GHz ≤ f < 22 GHz

add 0.5 dB to above values

22 GHz ≤ f < 26.5 GHz

add 1.0 dB to above values

26.5 GHz ≤ f ≤ 40 GHz

add 2.0 dB to above values

RF attenuation > 40 dB or
f ≥ 3.6 GHz, span ≥ 1 GHz


add 0.5 dB to above values

6

Only available in analyzer mode.

7

With option R&S

®

ESU-B24 the frequency range of the preamplifier is extended to the upper frequency limit of the instrument.