beautypg.com

Atec Rohde-Schwarz-ESU-40 User Manual

Page 18

background image

18 R&S

®

ESU EMI Test Receiver

Version 02.00, October 2007

RF attenuation = 0 dB, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW
preselection = ON, preamplifier = OFF
f < 10 kHz: RBW = 10 Hz, VBW = 30 Hz, trace average, sample detector,
sweep count = 20, mean marker
f ≥ 10 kHz: RBW = 1 kHz, VBW = 3 kHz, span = 0 Hz, sweep time = 50 ms, trace
average, sample detector, sweep count = 20, mean marker
all models

20 Hz

typ. <–90 dBm

100 Hz

<–110 dBm

1 kHz

<–120 dBm

10 kHz

<–130 dBm

100 kHz

<–130 dBm

1 MHz

<–140 dBm

10 MHz

<–153 dBm

R&S

®

ESU8:

20 MHz ≤ f < 2 GHz

<–155 dBm, typ. –158 dBm

2 GHz ≤ f < 2.5 GHz

<–151 dBm, typ. –154 dBm

2.5 GHz ≤ f < 3 GHz

<–148 dBm, typ. –151 dBm

3 GHz ≤ f < 3.6 GHz

<–142 dBm, typ. –145 dBm

3.6 GHz ≤ f < 5 GHz

<–148 dBm, typ. –151 dBm

5 GHz ≤ f < 6 GHz

<–147 dBm, typ. –150 dBm

6 GHz ≤ f ≤ 8 GHz

<–144 dBm, typ. –147 dBm

R&S

®

ESU26:

20 MHz ≤ f < 2 GHz

<–152 dBm, typ. –156 dBm

2 GHz ≤ f < 3 GHz

<–147 dBm, typ. –151 dBm

3 GHz ≤ f < 3.6 GHz

<–142 dBm, typ. –145 dBm

3.6 GHz ≤ f < 10 GHz

<–147 dBm, typ. –150 dBm

10 GHz ≤ f < 18 GHz

<–145 dBm, typ. –149 dBm

18 GHz ≤ f < 22 GHz

<–142 dBm, typ. –145 dBm

22 GHz ≤ f ≤ 26.5 GHz

<–140 dBm, typ. –143 dBm

R&S

®

ESU40:

20 MHz ≤ f < 2 GHz

<–152 dBm, typ. –156 dBm

2 GHz ≤ f < 3 GHz

<–147 dBm, typ. –151 dBm

3 GHz ≤ f < 3.6 GHz

<–142 dBm, typ. –145 dBm

3.6 GHz ≤ f < 10 GHz

<–147 dBm, typ. –150 dBm

10 GHz ≤ f < 18 GHz

<–145 dBm, typ. –149 dBm

18 GHz ≤ f < 22 GHz

<–142 dBm, typ. –145 dBm

22 GHz ≤ f < 26.5 GHz

<–140 dBm, typ. –143 dBm

26.5 GHz ≤ f < 32 GHz

<–135 dBm, typ. –138 dBm

32 GHz ≤ f ≤ 40 GHz

<–133 dBm, typ. –135 dBm